Clinical Interpretation of Urine Drug Tests

What Clinicians Need to Know About Urine Drug Screens

      Abstract

      Urine drug testing is frequently used in clinical, employment, educational, and legal settings and misinterpretation of test results can result in significant adverse consequences for the individual who is being tested. Advances in drug testing technology combined with a rise in the number of novel misused substances present challenges to clinicians to appropriately interpret urine drug test results. Authors searched PubMed and Google Scholar to identify published literature written in English between 1946 and 2016, using urine drug test, screen, false-positive, false-negative, abuse, and individual drugs of abuse as key words. Cited references were also used to identify the relevant literature. In this report, we review technical information related to detection methods of urine drug tests that are commonly used and provide an overview of false-positive/false-negative data for commonly misused substances in the following categories: cannabinoids, central nervous system (CNS) depressants, CNS stimulants, hallucinogens, designer drugs, and herbal drugs of abuse. We also present brief discussions of alcohol and tricyclic antidepressants as related to urine drug tests, for completeness. The goal of this review was to provide a useful tool for clinicians when interpreting urine drug test results and making appropriate clinical decisions on the basis of the information presented.

      Abbreviations and Acronyms:

      6-MAM (6-monoacetylmorphine), CNS (central nervous system), DMAA (dimethylamylamine), EtG (ethyglucuronide), ELISA (enzyme-linked immunosorbent assay), FDA (Food and Drug Administration), GC-MS (gas chromatography/mass spectrometry), LC-MS/MS (liquid chromatography/tandem mass spectrometry), MDPV (methylendioxyprovalerone), NSAID (nonsteroidal anti-inflammatory drug), POCT (point-of-care testing), PCP (phencyclidine), PPI (proton pump inhibitor), TCA (tricyclic antidepressant), THC (tetrahydrocannabinol), THCV (Δ9-tetrahydroccanbivarin), UDT (urine drug test)
      Article Highlights
      • Immunoassays have many weaknesses that can result in false-positive and false-negative results. Understanding how to interpret urine immunoassays (eg, cutoff values, detection times, and false-positive results) is vital when ordering.
      • All positive results on immunoassays need confirmatory testing (eg, gas chromatography/mass spectrometry).
      • Testing for designer drugs (eg, synthetic cathinones and cannabinoids) is challenging secondary to continual changes in synthetic compounds and increasing number of novel substances.
      There have been increased concerns regarding the nonmedical use of prescribed drugs and rising trends in illicit drug use and dependence. In 2014, it was estimated that 27 million Americans aged 12 years and older (representing 10.2% of the population) have used illicit drugs in the past month; this is compared with 7.9% in 2004.
      • Substance Abuse and Mental Health Services Administration
      Behavioral Health Trends in the United States: Results from the 2014 National Survey on Drug Use and Health.
      Urine drug testing is routinely used in clinical practice to rule out substance-induced disorders, confirm medication adherence, and identify substances in overdose situations. Employers and courts also perform drug tests to screen for illicit drug use. Despite the widespread use of urine drug tests (UDTs), there is little published information on how to correctly interpret the results of these tests. Incorrect interpretation of test results (false-positive or false-negative) can have significant consequences (eg, loss of job and incarceration). Unfortunately, there is evidence that there is a deficiency in clinician’s knowledge about accurate UDT interpretation.
      • Reisfield G.M.
      • Webb F.J.
      • Bertholf R.L.
      • Sloan P.A.
      • Wilson G.R.
      Family physicians’ proficiency in urine drug test interpretation.
      • Reisfield G.M.
      • Bertholf R.
      • Barkin R.L.
      • Webb F.
      • Wilson G.
      Urine drug test interpretation: what do physicians know?.
      Regular use of UDT did not correlate with increased knowledge; therefore, the need for clinician education may be widespread.
      The goal of this review was to provide an updated guide for clinicians that includes recent reports of agents that may cause false-positive results on common UDT immunoassays. We also expanded information on marijuana on the basis of recent legislative trends and included information on synthetic cathinones and cannabinoids. Our ultimate goal was to provide a concise reference that can be used in everyday practice by clinicians to accurately interpret UDT results that lead to appropriate therapeutic decisions.

      Literature Search

      Authors searched PubMed and Google Scholar to identify published literature between 1946 and 2016, using the following key words: urine drug test, screen, false-positive, false-negative, and abuse. In addition, individual drugs of abuse discussed in the article were also used as key words. For completeness, we also identified relevant cited references in the initially identified publications. Publications that discussed urinary testing of substances in humans or human samples only were selected.

      Methods of Drug Testing

      Drug testing can be completed on various biological specimens including urine, blood, hair, saliva, sweat, nails (toe and finger), and meconium. Urine is the most commonly obtained specimen for drug testing due to its noninvasive route and ease of sample collection. Both parent drug and metabolites may be detected in urine specimens and are usually in higher concentrations than in blood or serum samples. Drug detection times are longer in urine (ie, 1 day up to several weeks) than in blood or serum samples.
      • Substance Abuse and Mental Health Services Administration
      Clinical Drug Testing in Primary Care.
      There are 2 main types of UDTs, screening and confirmatory tests. Initial drug tests or screens are performed using immunoassay technology and are conducted in the laboratory or onsite with point-of-care testing (POCT). Immunoassays allow for a large number of specimen screens to be completed and provide relatively rapid results.
      • Armbruster D.A.
      • Schwarzhoff R.H.
      • Hubster E.C.
      • Liserio M.K.
      Enzyme immunoassay, kinetic microparticle immunoassay, radioimmunoassay, and fluorescence polarization immunoassay compared for drugs-of-abuse screening.
      Three main types of immunoassays are available: (1) enzyme-multiplied immunoassay technique, (2) enzyme-linked immunosorbent assay (ELISA), and (3) fluorescence polarization immunoassay. In general, immunoassays use antibodies to detect the presence of drug metabolites or classes of drug metabolites in the urine. Unfortunately, immunoassays will detect substances with similar characteristics, resulting in cross-reactivity leading to false-positive results.
      An increasing trend, especially in pain management clinics and with clinicians treating patients with substance use disorders, is POCT in the office setting. It allows for immediate results onsite, allowing the clinician to discuss results with the patient in real time. These POCTs should be cleared by the Food and Drug Administration (FDA) and are usually waived by Clinical Laboratory Improvement Amendments. Visual analysis of the test result provides interpretation of the outcomes. At times, results may be difficult to read (eg, faint color and uncertain color), leading to subjective interpretation.
      • Nichols J.H.
      • Christenson R.H.
      • Clarke W.
      • et al.
      National Academy of Clinical Biochemistry. Executive summary
      The National Academy of Clinical Biochemistry Laboratory Medicine Practice Guideline: evidence-based practice for point-of-care testing.
      In addition, POCT has the same limitations as laboratory-based immunoassays and results should be used only to screen for a substance. Consumers who purchase POCT kits are cautioned against interpreting any positive preliminary results and confirmatory testing by a professional is recommended.
      All initial testing conducted with immunoassays need to be considered presumptive, and clinicians need to use clinical judgment, patient history, and collaborative information to decide whether confirmatory testing is necessary for optimal patient care. Gas chromatography/mass spectrometry (GC-MS) is considered the criterion standard in confirmatory testing and can identify specific molecular structures and quantifies the amount of a drug or substance present in the sample.
      • Substance Abuse and Mental Health Services Administration
      Clinical Drug Testing in Primary Care.
      The GC-MS assessments must be conducted by highly trained personnel, are time-consuming and costly, and thus are reserved for confirming positive drug screens. Liquid chromatography/tandem mass spectrometry (LC-MS/MS) offers an alternative to GC-MS for confirmatory testing and may be more time-efficient. Confirmatory testing should always be conducted when making legal, forensic, academic, employment, or other decisions that have significant sequelae.

      Cutoff Levels

      Cutoff values for UDT define the concentrations needed to produce positive results for immunoassays and confirmation testing on GC-MS or LC-MS/MS. Cutoff levels were established to help minimize false-positive results especially in workplace drug testing (eg, passive inhalation of marijuana causing positive results; poppy seeds ingestion causing positive opiate results). Results lower than the established cutoff values are reported as negative. Therefore, a negative result does not indicate that a substance is not present, but that the concentration was lower than the established cutoff concentration. Table 1 displays the federal mandated cutoff levels for the workplace developed by the Department of Health and Human Services.
      Mandatory guidelines for federal workplace drug testing programs.
      Although clinicians should be aware of federal cutoff values for substances of abuse, they should recognize that the federal cutoff concentrations were established for use in the workplace in which higher cutoff concentrations may be necessary to avoid false-positive results.
      • Substance Abuse and Mental Health Services Administration
      Clinical Drug Testing in Primary Care.
      However, in medical practice, lower cutoff values may be necessary particularly when testing for medication adherence. Clinical laboratories may use cutoff levels that are different from federal guidelines; thus, it is important that practitioners are aware of the values when interpreting results. In addition, clinicians may need to request a lower cutoff value to be used to minimize false-negative results; however, this may increase the rate of false-positive results. Furthermore, cutoff values were established for the adult population. Lower cutoff values may be necessary for infants due to a more dilute urine.
      • Luzzi V.I.
      • Saunders A.N.
      • Koenig J.W.
      • et al.
      Analytic performance of immunoassays for drugs of abuse below established cutoff values.
      Urine osmolality tends to reach adult values after age 2 years.
      Table 1Federal Workplace Cutoff Values
      MDA = methylenedioxyamphetamine; MDMA = methylenedioxymethamphetamine; MDEA = methylenedioxyethylamphetamine.
      ,
      Mandatory guidelines for federal workplace drug testing programs.
      Initial test analyteInitial drug test level (immunoassay) (ng/mL)Confirmatory test analyteConfirmatory drug test level (GC-MS) (ng/mL)
      Marijuana metabolites50Delta-9-tetrahydrocannabinol-9-carboxylic acid15
      Cocaine metabolites150Benzoylecgonine100
      Opiate metabolites
       Codeine/morphine
      Morphine is the target analyte for codeine/morphine testing.
      2000Codeine/morphine2000
       6-Acetylmorphine106-Acetylmorphine10
      Phencyclidine25Phencyclidine25
      Amphetamine/methamphetamine
      Methamphetamine is the target analyte for amphetamine/methamphetamine testing.
      500Amphetamine250
      Methamphetamine
      Specimen must also contain amphetamine at a concentration greater than or equal to 100 ng/mL.
      250
      MDMA500MDMA250
      MDA250
      MDEA250
      a MDA = methylenedioxyamphetamine; MDMA = methylenedioxymethamphetamine; MDEA = methylenedioxyethylamphetamine.
      b Morphine is the target analyte for codeine/morphine testing.
      c Methamphetamine is the target analyte for amphetamine/methamphetamine testing.
      d Specimen must also contain amphetamine at a concentration greater than or equal to 100 ng/mL.

      Detection Times

      Detection time or window is the amount of time a drug can be detected in the urine and still produce a positive result. To evaluate detection times of a drug or substance, both drug characteristics and patient factors need to be considered. Drug characteristics include half-life, drug metabolites, drug interactions, dosing intervals, low versus high dosage, chronic versus occasional use, and time of last ingestion. Patient factors that also can affect detection times include body mass, pH of the urine, urine concentration, and renal or liver impairment. Table 2 reports standard detection times for drugs routinely detected in the urine.
      • Council on Scientific Affairs
      Scientific issues in drug testing.
      • Heit H.A.
      • Gourlay D.L.
      Urine drug testing in pain medicine.
      • Inaba D.S.
      • Cohen W.E.
      Uppers, Downers, All Arounders. Physical and Mental Effects of Psychoactive Drugs.
      • Moeller K.E.
      • Lee K.C.
      • Kissack J.C.
      Urine drug screening: practical guide for clinicians.
      NMS Labs Synthetic Cannabinoids ELISA FAQ. NMS Labs website.
      • Rosse R.B.
      • Deutsch L.H.
      • Deutsch S.I.
      Medical assessment and laboratory testing in psychiatry.
      Synthetic cannabinoid testing–urine. Frequently asked questions. Redwood Toxicology Laboratory website.
      • Verstraete A.G.
      Detection times of drugs of abuse in blood, urine, and oral fluid.
      • Woelfel J.A.
      Drug abuse urine tests: false-positive results.
      Table 2Approximate Drug Detection Time in the Urine
      • Council on Scientific Affairs
      Scientific issues in drug testing.
      • Heit H.A.
      • Gourlay D.L.
      Urine drug testing in pain medicine.
      • Inaba D.S.
      • Cohen W.E.
      Uppers, Downers, All Arounders. Physical and Mental Effects of Psychoactive Drugs.
      • Moeller K.E.
      • Lee K.C.
      • Kissack J.C.
      Urine drug screening: practical guide for clinicians.
      NMS Labs Synthetic Cannabinoids ELISA FAQ. NMS Labs website.
      • Rosse R.B.
      • Deutsch L.H.
      • Deutsch S.I.
      Medical assessment and laboratory testing in psychiatry.
      Synthetic cannabinoid testing–urine. Frequently asked questions. Redwood Toxicology Laboratory website.
      • Verstraete A.G.
      Detection times of drugs of abuse in blood, urine, and oral fluid.
      • Woelfel J.A.
      Drug abuse urine tests: false-positive results.
      Adapted from Mayo Clin Proc, with permission.
      • Moeller K.E.
      • Lee K.C.
      • Kissack J.C.
      Urine drug screening: practical guide for clinicians.
      DrugLength of time detected in urine
      Alcohol7-12 h
      Amphetamine48 h
      Methamphetamine48 h
      Barbiturate
       Short-acting (eg, pentobarbital)24 h
       Long-acting (eg, phenobarbital)3 wk
      Benzodiazepine
       Short-acting (eg, lorazepam)3 d
       Long-acting (eg, diazepam)30 d
      Cocaine metabolites2-4 d
      Marijuana
       Single use3 d
       Moderate use (4 times/wk)5-7 d
       Chronic use (daily)10-15 d
       Chronic heavy smoker>30 d
      Opioids
       Codeine48 h
       Heroin (morphine)48 h
       Hydromorphone2-4 d
      Methadone3 d
      Morphine48-72 h
      Oxycodone2-4 d
      Phencyclidine8 d
      Synthetic cannabinoids
       Single use72 h
       Chronic use>72 h
       Synthetic cathinoneVariable

      Evaluation of a Urine Sample

      People misusing drugs commonly use various methods (eg, adulteration, urine substitution, diluting urine) to avoid detection. A basic understanding of urine specimen characteristics is helpful to the clinician when evaluating drug screen results.
      Normal urine ranges from pale yellow to clear depending on its concentration. Specimens collected in the early morning have the highest concentration and therefore will contain higher levels of the drug.
      • Heit H.A.
      • Gourlay D.L.
      Urine drug testing in pain medicine.
      The temperature of the urine sample should be recorded within the first 4 minutes after collection and is usually between 90°F and 100°F.
      • Casavant M.J.
      Urine drug screening in adolescents.
      Urine specimen temperature may stay at 90.5°F for up to 15 minutes. Although urine pH fluctuates throughout the day, it generally ranges between 4.5 and 8. Specific gravity normally ranges between 1.002 and 1.030. In normal human urine, creatinine concentrations should be greater than 20 mg/dL. Urine specimens that are of unusual color or that are outside the normal parameters for human urine may be due to medications, foods, or disease states (diuretics, strict vegetarian diet, high state of hydration).
      • Hammett-Stabler C.A.
      • Pesce A.J.
      • Cannon D.J.
      Urine drug screening in the medical setting.
      It is imperative that documentation of these factors is included and be considered when the clinician is interpreting urine drug screen results.
      Adulteration or dilution of the urine specimen should be suspected if the pH is less than 3 or greater than 11 or the specific gravity is less than 1.002 or greater than 1.030.
      • Casavant M.J.
      Urine drug screening in adolescents.
      Urinary creatinine concentrations less than 20 mg/dL are indicative of dilute urine, whereas those less than 5 mg/dL combined with a specific gravity of less than 1.001 are not consistent with human urine.
      • Heit H.A.
      • Gourlay D.L.
      Urine drug testing in pain medicine.
      Urine specimens outside of these ranges are due to adulterations or dilution attempts. Urine specimens adulterated with soap may also produce excessive bubble formation that is long lasting.
      • Warner A.
      Interference of common household chemicals in immunoassay methods for drugs of abuse.
      If the urine specimen appears to be adulterated or diluted, the second specimen for evaluation should be collected under observation.
      Adulterants that have been used to mask a person’s use of a substance include household items such as table salt, laundry bleach, toilet bowl cleaner, vinegar, lemon juice, ammonia, or eye drops. Several select commercial adulterants containing glutaraldehyde (Clean X), sodium or potassium nitrite (Klear, Whizzies), pyridinium chlorochromate (Urine Luck), and peroxide/peroxidase (Stealth) are used to mask drug use.
      • Jaffee W.B.
      • Trucco E.
      • Levy S.
      • Weiss R.D.
      Is this urine really negative? A systematic review of tampering methods in urine drug screening and testing.
      Most household adulterants, except for eyedrops, can be detected by routine integrity (ie, temperature, pH, specific gravity) measurements.
      • Dasgupta A.
      The effects of adulterants and selected ingested compounds on drugs-of-abuse testing in urine.
      Commercial adulterants may mask the presence of drugs or their metabolites. Several dipstick tests (ie, AdultaCheck 4, AdultaCheck 6, Intect 7) are available for specimen integrity validation.
      • Dasgupta A.
      The effects of adulterants and selected ingested compounds on drugs-of-abuse testing in urine.

      Specific Drugs Tested in the Urine

      Determining which drug to test for in a UDT panel depends on the clinical setting. Most panels include the 5 drugs required by federal workplace guidelines, which include amphetamines, cocaine, marijuana, opiates, and phencyclidine.
      Mandatory guidelines for federal workplace drug testing programs.
      Benzodiazepines are commonly included in most UDTs. Clinicians working with patients with pain disorders should consider additional testing for semisynthetic (eg, oxycodone) and synthetic opioids (eg, fentanyl and methadone) (Table 3).
      • Substance Abuse and Mental Health Services Administration
      Clinical Drug Testing in Primary Care.
      Mass-spectrometry–based definitive laboratory testing should be considered once to twice per year on the basis of the risk of assessment.

      Langman LJ, Jannetto PJ, eds. Laboratory Medicine Practice Guidelines. Using Clinical Laboratory Tests to Monitor Drug Therapy in Pain Management Patients. Washington, DC: American Association for Clinical Chemistry, The National Academy of Clinical Biochemistry. https://www.aacc.org/∼/media/practice-guidelines/pain-management/rough-draft-pain-management-lmpg-v6aacc.pdf?la=en. Accessed October 16, 2016.

      Below, we discuss common drugs of abuse encountered in the clinical setting and common false-positives and false-negatives with each screening test (Table 4).
      • Moeller K.E.
      • Lee K.C.
      • Kissack J.C.
      Urine drug screening: practical guide for clinicians.
      • Woelfel J.A.
      Drug abuse urine tests: false-positive results.
      • Casavant M.J.
      Urine drug screening in adolescents.
      • Cody J.T.
      Precursor medications as a source of methamphetamine and/or amphetamine positive drug testing results.
      • Colbert D.L.
      Possible explanation for trimethobenzamide cross-reaction in immunoassays of amphetamine/methamphetamine.
      • Fenderson J.L.
      • Stratton A.N.
      • Domingo J.S.
      • Matthews G.O.
      • Tan C.D.
      Amphetamine positive urine toxicology screen secondary to atomoxetine.
      • Fucci N.
      False positive results for amphetamine in urine of a patient with diabetes mellitus.
      • Gilbert R.B.
      • Peng P.I.
      • Wong D.
      A labetalol metabolite with analytical characteristics resembling amphetamines.
      • Grinstead G.F.
      Ranitidine and high concentrations of phenylpropanolamine cross react in the EMIT monoclonal amphetamine/methamphetamine assay.
      • Jones R.
      • Klette K.
      • Kuhlman J.J.
      • et al.
      Trimethobenzamide cross-reacts in immunoassays of amphetamine/methamphetamine.
      • Kaplan J.
      • Shah P.
      • Faley B.
      • Siegel M.E.
      Case reports of aripiprazole causing false-positive urine amphetamine drug screens in children.
      • Kelly K.L.
      Ranitidine cross-reactivity in the EMIT d.a.u. Monoclonal Amphetamine/Methamphetamine Assay.
      • Levine B.S.
      • Caplan Y.H.
      Isometheptene cross reacts in the EMIT amphetamine assay.
      • Liu L.
      • Wheeler S.E.
      • Rymer J.A.
      • et al.
      Ranitidine interference with standard amphetamine immunoassay.
      • Manzi S.
      • Law T.
      • Shannon M.W.
      Methylphenidate produces a false-positive urine amphetamine screen.
      • Melanson S.E.
      • Lee-Lewandrowski E.
      • Griggs D.A.
      • Long W.H.
      • Flood J.G.
      Reduced interference by phenothiazines in amphetamine drug of abuse immunoassays.
      • Merigian K.S.
      • Browning R.
      • Kellerman A.
      Doxepin causing false-positive urine test for amphetamine.
      • Merigian K.S.
      • Browning R.G.
      Desipramine and amantadine causing false-positive urine test for amphetamine.
      • Nice A.
      • Maturen A.
      False-positive urine amphetamine screen with ritodrine.
      • Nixon A.L.
      • Long W.H.
      • Puopolo P.R.
      • Flood J.G.
      Bupropion metabolites produce false-positive urine amphetamine results.
      • Olsen K.M.
      • Gulliksen M.
      • Christophersen A.S.
      Metabolites of chlorpromazine and brompheniramine may cause false-positive urine amphetamine results with monoclonal EMIT d.a.u. immunoassay.
      • Poklis A.
      • Hall K.V.
      • Still J.
      • Binder S.R.
      Ranitidine interference with the monoclonal EMIT d.a.u. amphetamine/methamphetamine immunoassay.
      • Poklis A.
      • Moore K.A.
      Response of EMIT amphetamine immunoassays to urinary desoxyephedrine following Vicks inhaler use.
      • Roberge R.J.
      • Luellen J.R.
      • Reed S.
      False-positive amphetamine screen following a trazodone overdose.
      • Romberg R.W.
      • Needleman S.B.
      • Snyder J.J.
      • Greedan A.
      Methamphetamine and amphetamine derived from the metabolism of selegiline.
      • Stout P.R.
      • Klette K.L.
      • Horn C.K.
      Evaluation of ephedrine, pseudoephedrine and phenylpropanolamine concentrations in human urine samples and a comparison of the specificity of DRI amphetamines and Abuscreen online (KIMS) amphetamines screening immunoassays.
      • Vorce S.P.
      • Holler J.M.
      • Cawrse B.M.
      • Magluilo Jr., J.
      Dimethylamylamine: a drug causing positive immunoassay results for amphetamines.
      • Weintraub D.
      • Linder M.W.
      Amphetamine positive toxicology screen secondary to bupropion.
      • Yee L.M.
      • Wu D.
      False-positive amphetamine toxicology screen results in three pregnant women using labetalol.
      • Blank A.
      • Hellstern V.
      • Schuster D.
      • et al.
      Efavirenz treatment and false-positive results in benzodiazepine screening tests.
      • Fraser A.D.
      • Meatherall R.
      Comparative evaluation of five immunoassays for the analysis of alprazolam and triazolam metabolites in urine: effect of lowering the screening and GC-MS cut-off values.
      • Fraser A.D.
      • Meatherall R.
      Improved cross-reactivity to alpha OH triazolam in the BMC CEDIA DAU urine benzodiazepine assay.
      • Lum G.
      • Mushlin B.
      • Farney L.
      False-positive rates for the qualitative analysis of urine benzodiazepines and metabolites with the reformulated Abbott Multigent reagents.
      • Meatherall R.
      • Fraser A.D.
      Comparison of four immunoassays for the detection of lorazepam in urine.
      • Nasky K.M.
      • Cowan G.L.
      • Knittel D.R.
      False-positive urine screening for benzodiazepines: an association with sertraline? A two-year retrospective chart analysis.
      • Roder C.S.
      • Heinrich T.
      • Gehrig A.K.
      • Mikus G.
      Misleading results of screening for illicit drugs during efavirenz treatment.
      Tests for drugs of abuse.
      • Cotten S.W.
      • Duncan D.L.
      • Burch E.A.
      • Seashore C.J.
      • Hammett-Stabler C.A.
      Unexpected interference of baby wash products with a cannabinoid (THC) immunoassay.
      • Felton D.
      • Zitomersky N.
      • Manzi S.
      • Lightdale J.R.
      13-year-old girl with recurrent, episodic, persistent vomiting: out of the pot and into the fire.
      • Fraser A.D.
      • Meatherall R.
      Lack of interference by nabilone in the EMIT d.a.u. cannabinoid assay, Abbott TDx cannabinoid assay, and a sensitive TLC assay for delta 9-THC-carboxylic acid.
      • la Porte C.J.
      • Droste J.A.
      • Burger D.M.
      False-positive results in urine drug screening in healthy volunteers participating in phase 1 studies with efavirenz and rifampin.
      • Oosthuizen N.M.
      • Laurens J.B.
      Efavirenz interference in urine screening immunoassays for tetrahydrocannabinol.
      • Rollins D.E.
      • Jennison T.A.
      • Jones G.
      Investigation of interference by nonsteroidal anti-inflammatory drugs in urine tests for abused drugs.
      • Rossi S.
      • Yaksh T.
      • Bentley H.
      • van den Brande G.
      • Grant I.
      • Ellis R.
      Characterization of interference with 6 commercial delta9-tetrahydrocannabinol immunoassays by efavirenz (glucuronide) in urine.
      • Steinagle G.C.
      • Upfal M.
      Concentration of marijuana metabolites in the urine after ingestion of hemp seed tea.
      • De Giorgio F.
      • Rossi S.S.
      • Rainio J.
      • Chiarotti M.
      Cocaine found in a child’s hair due to environmental exposure?.
      • Hickey K.
      • Seliem R.
      • Shields J.
      • McKee A.
      • Nichols J.H.
      A positive drug test in the pain management patient: deception or herbal cross-reactivity?.
      • Mazor S.S.
      • Mycyk M.B.
      • Wills B.K.
      • Brace L.D.
      • Gussow L.
      • Erickson T.
      Coca tea consumption causes positive urine cocaine assay.
      • Baden L.R.
      • Horowitz G.
      • Jacoby H.
      • Eliopoulos G.M.
      Quinolones and false-positive urine screening for opiates by immunoassay technology.
      • Cooreman S.
      • Deprez C.
      • Martens F.
      • Van Bocxlaer J.
      • Croes K.
      A comprehensive LC-MS-based quantitative analysis of fentanyl-like drugs in plasma and urine.
      • Daher R.
      • Haidar J.H.
      • Al-Amin H.
      Rifampin interference with opiate immunoassays.
      • de Paula M.
      • Saiz L.C.
      • Gonzalez-Revalderia J.
      • Pascual T.
      • Alberola C.
      • Miravalles E.
      Rifampicin causes false-positive immunoassay results for urine opiates.
      • Herrera P.
      • Ortiz E.
      • Tena T.
      • Lora C.
      Presence of rifampicin in urine causes cross-reactivity with opiates using the KIMS method.
      • Kronstrand R.
      • Selden T.G.
      • Josefsson M.
      Analysis of buprenorphine, norbuprenorphine, and their glucuronides in urine by liquid chromatography-mass spectrometry.
      • Lichtenwalner M.R.
      • Mencken T.
      • Tully R.
      • Petosa M.
      False-positive immunochemical screen for methadone attributable to metabolites of verapamil.
      • Meatherall R.
      • Dai J.
      False-positive EMIT II opiates from ofloxacin.
      • Rogers S.C.
      • Pruitt C.W.
      • Crouch D.J.
      • Caravati E.M.
      Rapid urine drug screens: diphenhydramine and methadone cross-reactivity.
      • Straley C.M.
      • Cecil E.J.
      • Herriman M.P.
      Gatifloxacin interference with opiate urine drug screen.
      • Struempler R.E.
      Excretion of codeine and morphine following ingestion of poppy seeds.
      • Syed H.
      • Som S.
      • Khan N.
      • Faltas W.
      Doxylamine toxicity: seizure, rhabdomyolysis and false positive urine drug screen for methadone.
      • van As H.
      • Stolk L.M.
      Rifampicin cross-reacts with opiate immunoassay.
      • Vincent E.C.
      • Zebelman A.
      • Goodwin C.
      • Stephens M.M.
      Clinical inquiries: what common substances can cause false positives on urine screens for drugs of abuse?.
      • Wang G.
      • Huynh K.
      • Barhate R.
      • et al.
      Development of a homogeneous immunoassay for the detection of fentanyl in urine.
      • Zebelman A.M.
      • Troyer B.L.
      • Randall G.L.
      • Batjer J.D.
      Detection of morphine and codeine following consumption of poppy seeds.
      • Bond G.R.
      • Steele P.E.
      • Uges D.R.
      Massive venlafaxine overdose resulted in a false positive Abbott AxSYM urine immunoassay for phencyclidine.
      • Geraci M.J.
      • Peele J.
      • McCoy S.L.
      • Elias B.
      Phencyclidine false positive induced by lamotrigine (Lamictal(R)) on a rapid urine toxicology screen.
      • Gupta R.C.
      • Lu I.
      • Oei G.L.
      • Lundberg G.D.
      Determination of phencyclidine (PCP) in urine and illicit street drug samples.
      • Hull M.J.
      • Griggs D.
      • Knoepp S.M.
      • Smogorzewska A.
      • Nixon A.
      • Flood J.G.
      Postmortem urine immunoassay showing false-positive phencyclidine reactivity in a case of fatal tramadol overdose.
      • Khajawall A.M.
      • Simpson G.M.
      Critical interpretation of urinary phencyclidine monitoring.
      • Ly B.T.
      • Thornton S.L.
      • Buono C.
      • Stone J.A.
      • Wu A.H.
      False-positive urine phencyclidine immunoassay screen result caused by interference by tramadol and its metabolites.
      • Penders T.M.
      • Gestring R.E.
      • Vilensky D.A.
      Intoxication delirium following use of synthetic cathinone derivatives.
      • Rengarajan A.
      • Mullins M.E.
      How often do false-positive phencyclidine urine screens occur with use of common medications?.
      • Sena S.F.
      • Kazimi S.
      • Wu A.H.
      False-positive phencyclidine immunoassay results caused by venlafaxine and O-desmethylvenlafaxine.
      • Shannon M.
      Recent ketamine administration can produce a urine toxic screen which is falsely positive for phencyclidine.
      • Al-Mateen C.S.
      • Wolf II, C.E.
      Falsely elevated imipramine levels in a patient taking quetiapine.
      • Chattergoon D.S.
      • Verjee Z.
      • Anderson M.
      • et al.
      Carbamazepine interference with an immune assay for tricyclic antidepressants in plasma.
      • Dasgupta A.
      • Wells A.
      • Datta P.
      False-positive serum tricyclic antidepressant concentrations using fluorescence polarization immunoassay due to the presence of hydroxyzine and cetirizine.
      • Fleischman A.
      • Chiang V.W.
      Carbamazepine overdose recognized by a tricyclic antidepressant assay.
      • Matos M.E.
      • Burns M.M.
      • Shannon M.W.
      False-positive tricyclic antidepressant drug screen results leading to the diagnosis of carbamazepine intoxication.
      • Schussler J.M.
      • Juenke J.M.
      • Schussler I.
      Quetiapine and falsely elevated nortriptyline level.
      • Sloan K.L.
      • Haver V.M.
      • Saxon A.J.
      Quetiapine and false-positive urine drug testing for tricyclic antidepressants.
      • Sorisky A.
      • Watson D.C.
      Positive diphenhydramine interference in the EMIT-st assay for tricyclic antidepressants in serum.
      • Van Hoey N.M.
      Effect of cyclobenzaprine on tricyclic antidepressant assays.
      • Wians Jr., F.H.
      • Norton J.T.
      • Wirebaugh S.R.
      False-positive serum tricyclic antidepressant screen with cyproheptadine.
      • Yuan C.M.
      • Spandorfer P.R.
      • Miller S.L.
      • Henretig F.M.
      • Shaw L.M.
      Evaluation of tricyclic antidepressant false positivity in a pediatric case of cyproheptadine (periactin) overdose.
      The importance of confirmatory testing is emphasized to ensure an accurate and reliable UDT result.
      Table 3Classification of Opioids
      • Substance Abuse and Mental Health Services Administration
      Clinical Drug Testing in Primary Care.
      DerivationOpioid
      From opiumCodeine, morphine, opium, thebaine
      SemisyntheticBuprenorphine, dihydrocodeine, heroin, hydrocodone, hydromorphone, levorphanol, oxycodone, oxymorphone
      SyntheticFentanyl, meperidine, methadone, tramadol
      Table 4Summary of Agents Contributing to Results by Immunoassay
      MDMA = methylenedioxymethylamphetamine; MDPV = methylendioxyprovalerone; NSAID = nonsteroidal anti-inflammatory drug.
      Adapted from Mayo Clin Proc, with permission.
      • Moeller K.E.
      • Lee K.C.
      • Kissack J.C.
      Urine drug screening: practical guide for clinicians.
      SubstancePotential positives (includes true- and false-positives)Potential medications that may not be detected
      Alcohol
      Short-chain alcohols (eg, isopropyl alcohol)Not applicable
      Amphetamines
      • Cody J.T.
      Precursor medications as a source of methamphetamine and/or amphetamine positive drug testing results.
      • Colbert D.L.
      Possible explanation for trimethobenzamide cross-reaction in immunoassays of amphetamine/methamphetamine.
      • Fenderson J.L.
      • Stratton A.N.
      • Domingo J.S.
      • Matthews G.O.
      • Tan C.D.
      Amphetamine positive urine toxicology screen secondary to atomoxetine.
      • Fucci N.
      False positive results for amphetamine in urine of a patient with diabetes mellitus.
      • Gilbert R.B.
      • Peng P.I.
      • Wong D.
      A labetalol metabolite with analytical characteristics resembling amphetamines.
      • Grinstead G.F.
      Ranitidine and high concentrations of phenylpropanolamine cross react in the EMIT monoclonal amphetamine/methamphetamine assay.
      • Jones R.
      • Klette K.
      • Kuhlman J.J.
      • et al.
      Trimethobenzamide cross-reacts in immunoassays of amphetamine/methamphetamine.
      • Kaplan J.
      • Shah P.
      • Faley B.
      • Siegel M.E.
      Case reports of aripiprazole causing false-positive urine amphetamine drug screens in children.
      • Kelly K.L.
      Ranitidine cross-reactivity in the EMIT d.a.u. Monoclonal Amphetamine/Methamphetamine Assay.
      • Levine B.S.
      • Caplan Y.H.
      Isometheptene cross reacts in the EMIT amphetamine assay.
      • Liu L.
      • Wheeler S.E.
      • Rymer J.A.
      • et al.
      Ranitidine interference with standard amphetamine immunoassay.
      • Manzi S.
      • Law T.
      • Shannon M.W.
      Methylphenidate produces a false-positive urine amphetamine screen.
      • Melanson S.E.
      • Lee-Lewandrowski E.
      • Griggs D.A.
      • Long W.H.
      • Flood J.G.
      Reduced interference by phenothiazines in amphetamine drug of abuse immunoassays.
      • Merigian K.S.
      • Browning R.
      • Kellerman A.
      Doxepin causing false-positive urine test for amphetamine.
      • Merigian K.S.
      • Browning R.G.
      Desipramine and amantadine causing false-positive urine test for amphetamine.
      • Nice A.
      • Maturen A.
      False-positive urine amphetamine screen with ritodrine.
      • Nixon A.L.
      • Long W.H.
      • Puopolo P.R.
      • Flood J.G.
      Bupropion metabolites produce false-positive urine amphetamine results.
      • Olsen K.M.
      • Gulliksen M.
      • Christophersen A.S.
      Metabolites of chlorpromazine and brompheniramine may cause false-positive urine amphetamine results with monoclonal EMIT d.a.u. immunoassay.
      • Poklis A.
      • Hall K.V.
      • Still J.
      • Binder S.R.
      Ranitidine interference with the monoclonal EMIT d.a.u. amphetamine/methamphetamine immunoassay.
      • Poklis A.
      • Moore K.A.
      Response of EMIT amphetamine immunoassays to urinary desoxyephedrine following Vicks inhaler use.
      • Roberge R.J.
      • Luellen J.R.
      • Reed S.
      False-positive amphetamine screen following a trazodone overdose.
      • Romberg R.W.
      • Needleman S.B.
      • Snyder J.J.
      • Greedan A.
      Methamphetamine and amphetamine derived from the metabolism of selegiline.
      • Stout P.R.
      • Klette K.L.
      • Horn C.K.
      Evaluation of ephedrine, pseudoephedrine and phenylpropanolamine concentrations in human urine samples and a comparison of the specificity of DRI amphetamines and Abuscreen online (KIMS) amphetamines screening immunoassays.
      • Vorce S.P.
      • Holler J.M.
      • Cawrse B.M.
      • Magluilo Jr., J.
      Dimethylamylamine: a drug causing positive immunoassay results for amphetamines.
      • Weintraub D.
      • Linder M.W.
      Amphetamine positive toxicology screen secondary to bupropion.
      • Yee L.M.
      • Wu D.
      False-positive amphetamine toxicology screen results in three pregnant women using labetalol.
      l-Methamphetamine (Vick’s inhaler)
      Newer immunoassays have corrected the false-positive result for Vick’s inhaler.
      Not applicable
      l-Deprenyl
      Converts to l-methamphetamine and l-amphetamine.
      Amantadine
      Aripiprazole
      Atomoxetine
      Benzphetamine
      Bupropion
      Clobenzorex
      Approved in Mexico. Not approved in the United States.
      Chlorpromazine
      Desipramine
      Dextroamphetamine
      Dimethylamylamine
      Ephedrine
      Fenproporex
      Approved in Mexico. Not approved in the United States.
      Isometheptene
      Isoxsuprine
      Labetalol
      Metformin
      Methylphenidate
      Methamphetamine
      MDMA
      Phentermine
      Promethazine
      Pseudoephedrine
      Phenylephrine
      Phenylpropanolamine
      Ranitidine
      Ritodrine
      Selegiline
      Thioridazine
      Trazodone
      Trimipramine
      Trimethobenzamide
      Benzodiazepines
      • Blank A.
      • Hellstern V.
      • Schuster D.
      • et al.
      Efavirenz treatment and false-positive results in benzodiazepine screening tests.
      • Fraser A.D.
      • Meatherall R.
      Comparative evaluation of five immunoassays for the analysis of alprazolam and triazolam metabolites in urine: effect of lowering the screening and GC-MS cut-off values.
      • Fraser A.D.
      • Meatherall R.
      Improved cross-reactivity to alpha OH triazolam in the BMC CEDIA DAU urine benzodiazepine assay.
      • Lum G.
      • Mushlin B.
      • Farney L.
      False-positive rates for the qualitative analysis of urine benzodiazepines and metabolites with the reformulated Abbott Multigent reagents.
      • Meatherall R.
      • Fraser A.D.
      Comparison of four immunoassays for the detection of lorazepam in urine.
      • Nasky K.M.
      • Cowan G.L.
      • Knittel D.R.
      False-positive urine screening for benzodiazepines: an association with sertraline? A two-year retrospective chart analysis.
      • Roder C.S.
      • Heinrich T.
      • Gehrig A.K.
      • Mikus G.
      Misleading results of screening for illicit drugs during efavirenz treatment.
      EfavirenzAlprazolam
      OxaprozinClonazepam
      SertralineLorazepam
      Cannabinoids
      • Woelfel J.A.
      Drug abuse urine tests: false-positive results.
      Tests for drugs of abuse.
      • Cotten S.W.
      • Duncan D.L.
      • Burch E.A.
      • Seashore C.J.
      • Hammett-Stabler C.A.
      Unexpected interference of baby wash products with a cannabinoid (THC) immunoassay.
      • Felton D.
      • Zitomersky N.
      • Manzi S.
      • Lightdale J.R.
      13-year-old girl with recurrent, episodic, persistent vomiting: out of the pot and into the fire.
      • Fraser A.D.
      • Meatherall R.
      Lack of interference by nabilone in the EMIT d.a.u. cannabinoid assay, Abbott TDx cannabinoid assay, and a sensitive TLC assay for delta 9-THC-carboxylic acid.
      • la Porte C.J.
      • Droste J.A.
      • Burger D.M.
      False-positive results in urine drug screening in healthy volunteers participating in phase 1 studies with efavirenz and rifampin.
      • Oosthuizen N.M.
      • Laurens J.B.
      Efavirenz interference in urine screening immunoassays for tetrahydrocannabinol.
      • Rollins D.E.
      • Jennison T.A.
      • Jones G.
      Investigation of interference by nonsteroidal anti-inflammatory drugs in urine tests for abused drugs.
      • Rossi S.
      • Yaksh T.
      • Bentley H.
      • van den Brande G.
      • Grant I.
      • Ellis R.
      Characterization of interference with 6 commercial delta9-tetrahydrocannabinol immunoassays by efavirenz (glucuronide) in urine.
      • Steinagle G.C.
      • Upfal M.
      Concentration of marijuana metabolites in the urine after ingestion of hemp seed tea.
      Baby wash productsNabilone
      DronabinolSynthetic cannabinoids
      Efavirenz
      NSAIDs
      Proton pump inhibitors
      Cocaine
      • De Giorgio F.
      • Rossi S.S.
      • Rainio J.
      • Chiarotti M.
      Cocaine found in a child’s hair due to environmental exposure?.
      • Hickey K.
      • Seliem R.
      • Shields J.
      • McKee A.
      • Nichols J.H.
      A positive drug test in the pain management patient: deception or herbal cross-reactivity?.
      • Mazor S.S.
      • Mycyk M.B.
      • Wills B.K.
      • Brace L.D.
      • Gussow L.
      • Erickson T.
      Coca tea consumption causes positive urine cocaine assay.
      Coca leaf teaNot applicable
      Topical anesthetics containing cocaine
      Opioids/opiates/heroin
      • Woelfel J.A.
      Drug abuse urine tests: false-positive results.
      • Casavant M.J.
      Urine drug screening in adolescents.
      • Baden L.R.
      • Horowitz G.
      • Jacoby H.
      • Eliopoulos G.M.
      Quinolones and false-positive urine screening for opiates by immunoassay technology.
      • Cooreman S.
      • Deprez C.
      • Martens F.
      • Van Bocxlaer J.
      • Croes K.
      A comprehensive LC-MS-based quantitative analysis of fentanyl-like drugs in plasma and urine.
      • Daher R.
      • Haidar J.H.
      • Al-Amin H.
      Rifampin interference with opiate immunoassays.
      • de Paula M.
      • Saiz L.C.
      • Gonzalez-Revalderia J.
      • Pascual T.
      • Alberola C.
      • Miravalles E.
      Rifampicin causes false-positive immunoassay results for urine opiates.
      • Herrera P.
      • Ortiz E.
      • Tena T.
      • Lora C.
      Presence of rifampicin in urine causes cross-reactivity with opiates using the KIMS method.
      • Kronstrand R.
      • Selden T.G.
      • Josefsson M.
      Analysis of buprenorphine, norbuprenorphine, and their glucuronides in urine by liquid chromatography-mass spectrometry.
      • Lichtenwalner M.R.
      • Mencken T.
      • Tully R.
      • Petosa M.
      False-positive immunochemical screen for methadone attributable to metabolites of verapamil.
      • Meatherall R.
      • Dai J.
      False-positive EMIT II opiates from ofloxacin.
      • Rogers S.C.
      • Pruitt C.W.
      • Crouch D.J.
      • Caravati E.M.
      Rapid urine drug screens: diphenhydramine and methadone cross-reactivity.
      • Straley C.M.
      • Cecil E.J.
      • Herriman M.P.
      Gatifloxacin interference with opiate urine drug screen.
      • Struempler R.E.
      Excretion of codeine and morphine following ingestion of poppy seeds.
      • Syed H.
      • Som S.
      • Khan N.
      • Faltas W.
      Doxylamine toxicity: seizure, rhabdomyolysis and false positive urine drug screen for methadone.
      • van As H.
      • Stolk L.M.
      Rifampicin cross-reacts with opiate immunoassay.
      • Vincent E.C.
      • Zebelman A.
      • Goodwin C.
      • Stephens M.M.
      Clinical inquiries: what common substances can cause false positives on urine screens for drugs of abuse?.
      • Wang G.
      • Huynh K.
      • Barhate R.
      • et al.
      Development of a homogeneous immunoassay for the detection of fentanyl in urine.
      • Zebelman A.M.
      • Troyer B.L.
      • Randall G.L.
      • Batjer J.D.
      Detection of morphine and codeine following consumption of poppy seeds.
      DextromethorphanBuprenorphine
      Diphenhydramine
      Diphenhydramine, doxylamine, and verapamil (including metabolites) have been shown to cause positive results in methadone assays only.
      Fentanyl
      Doxylamine
      Diphenhydramine, doxylamine, and verapamil (including metabolites) have been shown to cause positive results in methadone assays only.
      Meperidine
      HeroinMethadone
      Opiates (codeine, hydromorphone, hydrocodone, morphine)Oxycodone
      Poppy seedsOxymorphone
      QuinineTramadol
      Quinolones
      Rifampin
      Verapamil and metabolites
      Diphenhydramine, doxylamine, and verapamil (including metabolites) have been shown to cause positive results in methadone assays only.
      Phencyclidine
      • Woelfel J.A.
      Drug abuse urine tests: false-positive results.
      • Bond G.R.
      • Steele P.E.
      • Uges D.R.
      Massive venlafaxine overdose resulted in a false positive Abbott AxSYM urine immunoassay for phencyclidine.
      • Geraci M.J.
      • Peele J.
      • McCoy S.L.
      • Elias B.
      Phencyclidine false positive induced by lamotrigine (Lamictal(R)) on a rapid urine toxicology screen.
      • Gupta R.C.
      • Lu I.
      • Oei G.L.
      • Lundberg G.D.
      Determination of phencyclidine (PCP) in urine and illicit street drug samples.
      • Hull M.J.
      • Griggs D.
      • Knoepp S.M.
      • Smogorzewska A.
      • Nixon A.
      • Flood J.G.
      Postmortem urine immunoassay showing false-positive phencyclidine reactivity in a case of fatal tramadol overdose.
      • Khajawall A.M.
      • Simpson G.M.
      Critical interpretation of urinary phencyclidine monitoring.
      • Ly B.T.
      • Thornton S.L.
      • Buono C.
      • Stone J.A.
      • Wu A.H.
      False-positive urine phencyclidine immunoassay screen result caused by interference by tramadol and its metabolites.
      • Penders T.M.
      • Gestring R.E.
      • Vilensky D.A.
      Intoxication delirium following use of synthetic cathinone derivatives.
      • Rengarajan A.
      • Mullins M.E.
      How often do false-positive phencyclidine urine screens occur with use of common medications?.
      • Sena S.F.
      • Kazimi S.
      • Wu A.H.
      False-positive phencyclidine immunoassay results caused by venlafaxine and O-desmethylvenlafaxine.
      • Shannon M.
      Recent ketamine administration can produce a urine toxic screen which is falsely positive for phencyclidine.
      DextromethorphanNot applicable
      Diphenhydramine
      Doxylamine
      Ibuprofen
      Imipramine
      Ketamine
      Lamotrigine
      MDPV
      Meperidine
      Mesoridazine
      Thioridazine
      Tramadol
      Venlafaxine, O-desmethylvenlafaxine
      Tricyclic antidepressants
      • Al-Mateen C.S.
      • Wolf II, C.E.
      Falsely elevated imipramine levels in a patient taking quetiapine.
      • Chattergoon D.S.
      • Verjee Z.
      • Anderson M.
      • et al.
      Carbamazepine interference with an immune assay for tricyclic antidepressants in plasma.
      • Dasgupta A.
      • Wells A.
      • Datta P.
      False-positive serum tricyclic antidepressant concentrations using fluorescence polarization immunoassay due to the presence of hydroxyzine and cetirizine.
      • Fleischman A.
      • Chiang V.W.
      Carbamazepine overdose recognized by a tricyclic antidepressant assay.
      • Matos M.E.
      • Burns M.M.
      • Shannon M.W.
      False-positive tricyclic antidepressant drug screen results leading to the diagnosis of carbamazepine intoxication.
      • Schussler J.M.
      • Juenke J.M.
      • Schussler I.
      Quetiapine and falsely elevated nortriptyline level.
      • Sloan K.L.
      • Haver V.M.
      • Saxon A.J.
      Quetiapine and false-positive urine drug testing for tricyclic antidepressants.
      • Sorisky A.
      • Watson D.C.
      Positive diphenhydramine interference in the EMIT-st assay for tricyclic antidepressants in serum.
      • Van Hoey N.M.
      Effect of cyclobenzaprine on tricyclic antidepressant assays.
      • Wians Jr., F.H.
      • Norton J.T.
      • Wirebaugh S.R.
      False-positive serum tricyclic antidepressant screen with cyproheptadine.
      • Yuan C.M.
      • Spandorfer P.R.
      • Miller S.L.
      • Henretig F.M.
      • Shaw L.M.
      Evaluation of tricyclic antidepressant false positivity in a pediatric case of cyproheptadine (periactin) overdose.
      Carbamazepine
      Reports of false-positive results occurred in serum only.
      Not applicable
      Cyclobenzaprine
      Cyproheptadine
      Reports of false-positive results occurred in serum only.
      Diphenhydramine
      Reports of false-positive results occurred in serum only.
      Hydroxyzine
      Reports of false-positive results occurred in serum only.
      Quetiapine
      Synthetic cannabinoids
      LamotrigineNot applicable
      a MDMA = methylenedioxymethylamphetamine; MDPV = methylendioxyprovalerone; NSAID = nonsteroidal anti-inflammatory drug.
      b Newer immunoassays have corrected the false-positive result for Vick’s inhaler.
      c Converts to l-methamphetamine and l-amphetamine.
      d Approved in Mexico. Not approved in the United States.
      e Diphenhydramine, doxylamine, and verapamil (including metabolites) have been shown to cause positive results in methadone assays only.
      f Reports of false-positive results occurred in serum only.

      Cannabinoids

      Cannabis or marijuana generally refers to any part of the Cannabis plant and has been used throughout history for textiles, fuels, and medicines and for its euphoric effects.
      • Inaba D.S.
      • Cohen W.E.
      Uppers, Downers, All Arounders. Physical and Mental Effects of Psychoactive Drugs.
      The Cannabis plant contains approximately 460 active chemicals with more than 60 chemicals classified as cannabinoids. Delta-9-tetrahydrocannabinol (THC) is considered the primary active chemical responsible for marijuana’s medicinal and psychoactive effects.
      Currently, marijuana is the most widely used “illicit” substance in the United States, with almost 20 million Americans 12 years or older using marijuana in 2013.
      • Substance Abuse and Mental Health Services Administration
      Results from the 2013 National Survey on Drug Use and Health: Summary of National Findings.
      Smoking or inhaling marijuana through cigarettes, cigars, water pipes, or vaporization is the most common route of administration primarily due to its rapid effects and ability to deliver high concentrations of the drug into the bloodstream.
      • National Institute on Drug Abuse
      Marijuana.
      Some users prefer the oral route of administration by mixing marijuana’s oil base extract (hash oil) into common foods such as desserts, candies, or sodas.
      Although illegal by the federal government, as of November 2016, 28 states plus the District of Columbia have approved marijuana use for medical purposes, and 8 states including the District of Columbia have approved marijuana for recreational use.
      28 legal medical marijuana states and DC: laws, fees, and possession limits.
      With state legalizations, it is important that clinicians inquire about medical and recreational marijuana use when ordering a drug screen to help with interpretation. It is important for users of medical or recreational marijuana to be aware that although approved by their state government, other entities (eg, federal systems, workplace, criminal justice systems, and schools) may still require a negative drug test result for marijuana. Furthermore, clinicians need to consider unintentional ingestion of marijuana especially in the presence of unexplainable neurologic conditions and food-borne illness. Both adults and children are susceptible to accidental ingestion of marijuana, especially through unlabeled food products.
      • Centers for Disease Control and Prevention
      Inadvertent ingestion of marijuana—Los Angeles, California, 2009.
      Children may experience more profound effects due to the edibles containing unverified dosages, and adults who have never used illicit drugs may experience more adverse effects. Reports of accidental ingestions have increased markedly since the legalization of marijuana in various states, and clinicians need to consider ordering a UDT for THC when necessary.
      Urine drug testing for marijuana is based on THC’s main metabolite 11-nor-delta-9-tetrahydrocannabinol-9-carboxylic acid.
      Mandatory guidelines for federal workplace drug testing programs.
      • Ellis Jr., G.M.
      • Mann M.A.
      • Judson B.A.
      • Schramm N.T.
      • Tashchian A.
      Excretion patterns of cannabinoid metabolites after last use in a group of chronic users.
      Initial testing through immunoassay is sensitive to several THC metabolites and the federal cutoff level is 50 ng/mL although some laboratories may use a lower cutoff level of 20 ng/mL.
      Mandatory guidelines for federal workplace drug testing programs.
      Confirmation testing via GC-MS or LC-MS/MS is specific for 9-tetrahydrocannabinol-9-carboxylic acid, allowing for a lower federal cutoff concentration of 15 ng/mL.
      Mandatory guidelines for federal workplace drug testing programs.
      • Ellis Jr., G.M.
      • Mann M.A.
      • Judson B.A.
      • Schramm N.T.
      • Tashchian A.
      Excretion patterns of cannabinoid metabolites after last use in a group of chronic users.
      Estimating the detection time for marijuana in the urine is multifaceted. Factors that influence detection of marijuana include route of administration, dosage and potency of marijuana, frequency of use, body mass, and one’s metabolic rate. Cannabinoids are highly lipophilic and are extensively stored in lipid compartments throughout the body. Chronic use of marijuana will result in accumulation of THC in fatty tissues, resulting in slow elimination rates of marijuana metabolites.
      • Ellis Jr., G.M.
      • Mann M.A.
      • Judson B.A.
      • Schramm N.T.
      • Tashchian A.
      Excretion patterns of cannabinoid metabolites after last use in a group of chronic users.
      Detection of marijuana can occur in the urine for greater than 30 days after cessation among chronic users,
      • Ellis Jr., G.M.
      • Mann M.A.
      • Judson B.A.
      • Schramm N.T.
      • Tashchian A.
      Excretion patterns of cannabinoid metabolites after last use in a group of chronic users.
      • Dackis C.A.
      • Pottash A.L.
      • Annitto W.
      • Gold M.S.
      Persistence of urinary marijuana levels after supervised abstinence.
      whereas single exposure to marijuana in nonusers typically can be detected in the urine only up to 72 hours.
      • Hollister L.E.
      • Kanter S.L.
      Laboratory verification of “heavy” and “light” users of cannabis.
      A practical challenge with UDT for marijuana is determining acute versus chronic marijuana use. Researchers have looked at quantifying the glucuronide conjugates of THC and 11-OH-THC (using Escherichia coli β-glucosidase hydrolysis) as biomarkers for recent (<8 hours) marijuana consumption.
      • Lowe R.H.
      • Abraham T.T.
      • Darwin W.D.
      • Herning R.
      • Cadet J.L.
      • Huestis M.A.
      Extended urinary delta9-tetrahydrocannabinol excretion in chronic cannabis users precludes use as a biomarker of new drug exposure.
      • Manno J.E.
      • Manno B.R.
      • Kemp P.M.
      • et al.
      Temporal indication of marijuana use can be estimated from plasma and urine concentrations of delta9-tetrahydrocannabinol, 11-hydroxy-delta9-tetrahydrocannabinol, and 11-nor-delta9-tetrahydrocannabinol-9-carboxylic acid.
      However, Lowe et al
      • Lowe R.H.
      • Abraham T.T.
      • Darwin W.D.
      • Herning R.
      • Cadet J.L.
      • Huestis M.A.
      Extended urinary delta9-tetrahydrocannabinol excretion in chronic cannabis users precludes use as a biomarker of new drug exposure.
      found concentrations in the urine up to 24 days after cessation in a chronic heavy user, refuting the effectiveness of these biomarkers. With respect to driving under the influence, most states rely on blood levels to determine impairment.
      • Lacey J.
      • Brainard K.
      • Snitow S.
      Drug Per Se Laws: A Review of Their Use in States.
      However, blood concentrations can rapidly decline within the first hour because of rapid distribution into fat stores and first-pass hepatic metabolism.
      • Huestis M.A.
      • Henningfield J.E.
      • Cone E.J.
      Blood cannabinoids, I: absorption of THC and formation of 11-OH-THC and THCCOOH during and after smoking marijuana.
      In addition, Bergamaschi et al
      • Bergamaschi M.M.
      • Karschner E.L.
      • Goodwin R.S.
      • et al.
      Impact of prolonged cannabinoid excretion in chronic daily cannabis smokers’ blood on per se drugged driving laws.
      found detectable THC concentrations in the blood after 30 days in 5 patients.
      There are 2 FDA-approved prescription medication forms of THC. Dronabinol, a synthetic version of THC, and nabilone, a synthetic cannabinoid similar to THC, are indicated for chemotherapy-induced emesis and anorexia in patients with AIDS (dronabinol only).
      Dronabinol will test positive for THC on UDTs, whereas nabilone tests negative for THC due to its distinct metabolites.
      • Fraser A.D.
      • Meatherall R.
      Lack of interference by nabilone in the EMIT d.a.u. cannabinoid assay, Abbott TDx cannabinoid assay, and a sensitive TLC assay for delta 9-THC-carboxylic acid.
      A challenge in patients receiving dronabinol is the inability to distinguish dronabinol from plant THC through confirmatory tests. Levin et al
      • Levin F.R.
      • Mariani J.J.
      • Brooks D.J.
      • Xie S.
      • Murray K.A.
      Delta9-tetrahydrocannabivarin testing may not have the sensitivity to detect marijuana use among individuals ingesting dronabinol.
      conducted a study to determine whether testing for Δ9-tetrahydroccanbivarin (THCV), a plant cannabinoid, in the urine would help distinguish the use of illicit plant marijuana use from oral dronabinol use. However, only 50% of participants who used cannabis heavily (≥5 times per week) tested positive for THCV. The authors concluded that this test was not sensitive enough to test for either the presence or absence of THCV likely owing to variable strains of cannabis.
      Medications reported to cross-react with cannabinoid immunoassays include proton pump inhibitors (PPIs),
      • Felton D.
      • Zitomersky N.
      • Manzi S.
      • Lightdale J.R.
      13-year-old girl with recurrent, episodic, persistent vomiting: out of the pot and into the fire.
      nonsteroidal anti-inflammatory drugs (NSAIDs),
      • Rollins D.E.
      • Jennison T.A.
      • Jones G.
      Investigation of interference by nonsteroidal anti-inflammatory drugs in urine tests for abused drugs.
      and efavirenz.
      • Oosthuizen N.M.
      • Laurens J.B.
      Efavirenz interference in urine screening immunoassays for tetrahydrocannabinol.
      • Rossi S.
      • Yaksh T.
      • Bentley H.
      • van den Brande G.
      • Grant I.
      • Ellis R.
      Characterization of interference with 6 commercial delta9-tetrahydrocannabinol immunoassays by efavirenz (glucuronide) in urine.
      Literature describing the interference of PPIs with UDT is limited to 1 case report of pantoprazole and pantoprazole’s package insert.
      • Felton D.
      • Zitomersky N.
      • Manzi S.
      • Lightdale J.R.
      13-year-old girl with recurrent, episodic, persistent vomiting: out of the pot and into the fire.
      The mechanism for pantoprazole’s interference with marijuana’s UDT is unknown and it is unclear whether this is a class effect or limited only to pantoprazole. Prescribing information of other PPIs does not report this interference.
      With respect to NSAIDs, Rollins et al
      • Rollins D.E.
      • Jennison T.A.
      • Jones G.
      Investigation of interference by nonsteroidal anti-inflammatory drugs in urine tests for abused drugs.
      found that only 2 samples out of 510 samples produced false-positive results for cannabis on immunoassay, 1 in a patient who took a single daily dose of 1200 mg of ibuprofen and 1 in a chronic naproxen user. NSAID interference appears to be rare; however, secondary confirmation is warranted in patients using NSAIDs with unexplained THC results on immunoassay.
      The cross-reactivity of efavirenz, a nonnucleoside reverse transcriptase inhibitor, on UDT for marijuana has been well documented.
      • Oosthuizen N.M.
      • Laurens J.B.
      Efavirenz interference in urine screening immunoassays for tetrahydrocannabinol.
      • Rossi S.
      • Yaksh T.
      • Bentley H.
      • van den Brande G.
      • Grant I.
      • Ellis R.
      Characterization of interference with 6 commercial delta9-tetrahydrocannabinol immunoassays by efavirenz (glucuronide) in urine.
      The glucuronide metabolite (EFV-8-ether glucuronide) has been attributed to causing the false-positive result.
      Surface contaminants with urine collections have also been shown to cause false-positive results in UDTs in newborns. Because of an increase in false-positive rates for THC UDTs in newborns, Cotten et al
      • Cotten S.W.
      • Duncan D.L.
      • Burch E.A.
      • Seashore C.J.
      • Hammett-Stabler C.A.
      Unexpected interference of baby wash products with a cannabinoid (THC) immunoassay.
      investigated several commercial products and materials (eg, baby wash, wipes, diapers, and urine collection bags) to determine whether cross-reactivity was present. Several baby wash products produced a dose-dependent response on THC immunoassays, with many testing positive using a cutoff of 20 ng/mL but none reach the standard cutoff level of 50 ng/mL. It was discovered that nurses used different techniques to clean newborns before and during sample collection. This study highlights the importance of surface contaminants especially in the collection and analysis of urine in newborns.
      A rising concern, especially with state approval of recreational marijuana use, is whether second-hand exposure to marijuana can result in positive drug screening. Several studies were conducted in the 1980s evaluating whether passive inhalation of marijuana would test positive on cannabis urine assays.
      • Mulé S.J.
      • Lomax P.
      • Gross S.J.
      Active and realistic passive marijuana exposure tested by three immunoassays and GC/MS in urine.
      • Perez-Reyes M.
      • Di Guiseppi S.
      • Mason A.P.
      • Davis K.H.
      Passive inhalation of marihuana smoke and urinary excretion of cannabinoids.
      • Cone E.J.
      • Johnson R.E.
      • Darwin W.D.
      • et al.
      Passive inhalation of marijuana smoke: urinalysis and room air levels of delta-9-tetrahydrocannabinol.
      Most of these studies found detectable urine concentrations of THC’s metabolites significantly below standard cutoff values. Since these studies were conducted, the potency of marijuana has significantly increased. In the 1980s, the potency of THC confiscated by law enforcement was around 3% whereas in 2014 the potency was approximately 12%.
      • ElSohly M.A.
      • Mehmedic Z.
      • Foster S.
      • Gon C.
      • Chandra S.
      • Church J.C.
      Changes in cannabis potency over the last 2 decades (1995-2014): analysis of current data in the United States.
      • ElSohly M.A.
      • Ross S.A.
      • Mehmedic Z.
      • Arafat R.
      • Yi B.
      • Banahan III, B.F.
      Potency trends of delta9-THC and other cannabinoids in confiscated marijuana from 1980-1997.
      With the rise in THC potency, Cone et al
      • Cone E.J.
      • Bigelow G.E.
      • Herrmann E.S.
      • et al.
      Non-smoker exposure to secondhand cannabis smoke, I: urine screening and confirmation results.
      evaluated the effects of passive inhalation with high-potency THC (up to 11.3%). The study placed 6 nonsmokers in a small room for 1 hour with smokers under the following conditions: (1) without air ventilation with participants actively smoking marijuana cigarettes containing 5.3% THC, (2) without air ventilation with participants actively smoking marijuana cigarettes containing 11.3% THC, and (3) with active air ventilation with participants actively smoking marijuana cigarettes containing 11.3% THC.
      • Cone E.J.
      • Bigelow G.E.
      • Herrmann E.S.
      • et al.
      Non-smoker exposure to secondhand cannabis smoke, I: urine screening and confirmation results.
      • Herrmann E.S.
      • Cone E.J.
      • Mitchell J.M.
      • et al.
      Non-smoker exposure to secondhand cannabis smoke, II: effect of room ventilation on the physiological, subjective, and behavioral/cognitive effects.
      None of the participants tested positive with an immunoassay (ELISA) cutoff level of more than 20 ng/mL in the room with ventilation. In rooms without ventilation, multiple immunoassays tested positive when using a 20 ng/mL cutoff value and 1 tested positive at the 50 ng/mL cutoff value (condition 2). Detection times to produce a positive screen (ELISA >20 ng/mL) ranged from 2 to 22 hours postexposure. Although 1 nonsmoker met the federal cutoff concentration and many with lower cutoff values, detection time was short and it was under harsh conditions (no ventilation) in which someone would be aware they were heavily exposed to second-hand smoke.

      Central Nervous System Depressants

      Opioids

      “Opioid” is the term to describe all compounds that work at the opioid receptors in the central nervous system (CNS) and peripheral tissues. Opioids are primarily used for their analgesic properties, although they also have antitussive or antidiarrheal effects. Common prescription opioid medications include morphine, hydrocodone, hydromorphone, oxycodone, fentanyl, methadone, and tramadol, while heroin is an illicit agent. The term “opiates” is used only to describe morphine and codeine, which are naturally derived from the opium poppy seed.
      • Trescot A.M.
      • Datta S.
      • Lee M.
      • Hansen H.
      Opioid pharmacology.
      Table 3 categorizes the opioid compounds according to sources of derivation.
      • Substance Abuse and Mental Health Services Administration
      Clinical Drug Testing in Primary Care.
      Opioid use has drastically increased in the past 10 to 15 years, and subsequently opioid misuse and abuse are also on the rise.
      • Peppin J.F.
      • Passik S.D.
      • Couto J.E.
      • et al.
      Recommendations for urine drug monitoring as a component of opioid therapy in the treatment of chronic pain.
      All prescription opioids have the potential for abuse and are Schedule II medications except tramadol, which recently went from unscheduled status to Schedule IV.
      Drug Enforcement Administration, Department of Justice
      Schedule of controlled substances: placement of tramadol into schedule IV: final rule.
      With the recent rescheduling of hydrocodone products from Schedule III to II, it is expected that there may be an increase in tramadol prescriptions due to ease of prescribing Schedule IV medications compared with Schedule II medications.
      Drug Enforcement Administration
      Department of Justice. Schedules of controlled substances: rescheduling of hydrocodone combination products from schedule III to schedule II.
      Urine drug testing is one of many tools for safe prescribing of opioids with appropriate assessment and monitoring.
      • Gourlay D.L.
      • Heit H.A.
      • Almahrezi A.
      Universal precautions in pain medicine: a rational approach to the treatment of chronic pain.
      • Passik S.D.
      • Kirsh K.L.
      • Whitcomb L.
      • et al.
      A new tool to assess and document pain outcomes in chronic pain patients receiving opioid therapy.
      It is important for clinicians to be aware that UDTs may not detect all opioid drugs equally, and it is vital that clinicians ordering UDT for opioids know how to interpret results, are familiar with which agents their laboratory tests for, and understands opium metabolism (Figure 1).
      • Peppin J.F.
      • Passik S.D.
      • Couto J.E.
      • et al.
      Recommendations for urine drug monitoring as a component of opioid therapy in the treatment of chronic pain.
      • Smith H.S.
      Opioid metabolism.
      Most conventional immunoassays use morphine as a single calibrator drug to set the threshold for distinguishing a “positive” or “negative” test result. Because cross-reactivity of antibodies between morphine and other opiates such as oxycodone, hydrocodone, hydromorphone, and oxymorphone is low, there may be a risk of false-negative results.
      • Smith M.L.
      • Hughes R.O.
      • Levine B.
      • Dickerson S.
      • Darwin W.D.
      • Cone E.J.
      Forensic drug testing for opiates, VI: urine testing for hydromorphone, hydrocodone, oxymorphone, and oxycodone with commercial opiate immunoassays and gas chromatography-mass spectrometry.
      More advanced immunoassays or LC-MS/MS show higher specificity that can maximize detection for those agents.
      • Smith M.L.
      • Hughes R.O.
      • Levine B.
      • Dickerson S.
      • Darwin W.D.
      • Cone E.J.
      Forensic drug testing for opiates, VI: urine testing for hydromorphone, hydrocodone, oxymorphone, and oxycodone with commercial opiate immunoassays and gas chromatography-mass spectrometry.
      Fentanyl, methadone, and buprenorphine have distinct differences in chemical structure compared with morphine; thus, there is no reactivity in commonly marketed morphine-specific immunoassays
      • Wang G.
      • Huynh K.
      • Barhate R.
      • et al.
      Development of a homogeneous immunoassay for the detection of fentanyl in urine.
      • Melanson S.E.
      • Snyder M.L.
      • Jarolim P.
      • Flood J.G.
      A new highly specific buprenorphine immunoassay for monitoring buprenorphine compliance and abuse.
      and these require immunoassays that are specific for these compounds or LC-MS/MS.
      • Cooreman S.
      • Deprez C.
      • Martens F.
      • Van Bocxlaer J.
      • Croes K.
      A comprehensive LC-MS-based quantitative analysis of fentanyl-like drugs in plasma and urine.
      • Kronstrand R.
      • Selden T.G.
      • Josefsson M.
      Analysis of buprenorphine, norbuprenorphine, and their glucuronides in urine by liquid chromatography-mass spectrometry.
      In addition, some laboratories do not routinely test for semisynthetic or synthetic medication (see Table 3) in a standard opioid UDT unless specially requested. Clinicians must have an adequate understanding of their institution’s laboratory immunoassay capabilities and/or the option for LC-MS/MS before using UDT.
      Figure thumbnail gr1
      Figure 1Metabolism of opioids.
      • Peppin J.F.
      • Passik S.D.
      • Couto J.E.
      • et al.
      Recommendations for urine drug monitoring as a component of opioid therapy in the treatment of chronic pain.
      • Smith H.S.
      Opioid metabolism.
      aMorphine is metabolized to hydromorphone in very small amounts.
      Another clinical limitation with UDT for federal and Department of Transportation–regulated industry employees is the federal cutoff level of 2000 ng/mL with additional testing for heroin metabolite 6-monoacethyl-morphine with a cutoff of 10 ng/mL use.
      Mandatory guidelines for federal workplace drug testing programs.
      The cutoff level for opiate (eg, morphine and codeine) testing was raised from 300 ng/mL to 2000 ng/mL of morphine in 1998 in efforts to limit the large number of morphine and/or codeine positive results from poppy seed ingestion or routine prescription opiate use when screening for heroin abuse.
      • Struempler R.E.
      Excretion of codeine and morphine following ingestion of poppy seeds.
      • Zebelman A.M.
      • Troyer B.L.
      • Randall G.L.
      • Batjer J.D.
      Detection of morphine and codeine following consumption of poppy seeds.
      Unfortunately, using this high workplace drug testing cutoff level can result in negative test results, making it difficult for clinicians to interpret recent opioid use especially when testing for synthetic and semisynthetic opioids.
      • Paul B.D.
      • Shimomura E.T.
      • Smith M.L.
      A practical approach to determine cutoff concentrations for opiate testing with simultaneous detection of codeine, morphine, and 6-acetylmorphine in urine.
      Clinicians who commonly prescribe opioid medications for chronic pain and use UDT for compliance monitoring and abuse detection may need to use the lower threshold of 300 ng/mL. As clinicians, it is important that one is aware of their laboratory’s cutoff value for opioids and when necessary may need to request additional testing at a lower cutoff. The most common reasons for opioid false-negative results are using incorrect testing for a specific opioid or there is insufficient concentration of opioid in the urine.
      • Keary C.J.
      • Wang Y.
      • Moran J.R.
      • Zayas L.V.
      • Stern T.A.
      Toxicologic testing for opiates: understanding false-positive and false-negative test results.
      A few nonopioid agents have been shown to cause false-positive results for opiates and are reported in Table 4. Quinolones, which are commonly prescribed antiinfectives, are widely reported to interfere with opiate immunoassays.
      • Meatherall R.
      • Dai J.
      False-positive EMIT II opiates from ofloxacin.
      • Straley C.M.
      • Cecil E.J.
      • Herriman M.P.
      Gatifloxacin interference with opiate urine drug screen.
      Meatherall and Dai
      • Meatherall R.
      • Dai J.
      False-positive EMIT II opiates from ofloxacin.
      evaluated ofloxacin, norfloxacin, and ciprofloxacin for cross-reactivity on the enzyme-multiplied immunoassay technique II opiate immunoassay using a morphine threshold of 300 ng/mL. Ofloxacin was found to produce positive results, whereas norfloxacin and ciprofloxacin did not elicit positive results. Gatifloxacin also was found in a case report to provide a positive finding for opiates using the 2000 ng/mL cutoff level.
      • Straley C.M.
      • Cecil E.J.
      • Herriman M.P.
      Gatifloxacin interference with opiate urine drug screen.
      Rifampin or rifampicin also has caused false-positive results with opiate immunoassays.
      • Daher R.
      • Haidar J.H.
      • Al-Amin H.
      Rifampin interference with opiate immunoassays.
      • de Paula M.
      • Saiz L.C.
      • Gonzalez-Revalderia J.
      • Pascual T.
      • Alberola C.
      • Miravalles E.
      Rifampicin causes false-positive immunoassay results for urine opiates.
      • Herrera P.
      • Ortiz E.
      • Tena T.
      • Lora C.
      Presence of rifampicin in urine causes cross-reactivity with opiates using the KIMS method.
      • van As H.
      • Stolk L.M.
      Rifampicin cross-reacts with opiate immunoassay.
      To understand opiate UDT, a proper understanding of specific opioid metabolism is essential. Studies have shown that clinicians struggle with interpreting opioid UDT results, which may be due to lack of understanding of opioid metabolism.
      • Reisfield G.M.
      • Webb F.J.
      • Bertholf R.L.
      • Sloan P.A.
      • Wilson G.R.
      Family physicians’ proficiency in urine drug test interpretation.
      • Reisfield G.M.
      • Bertholf R.
      • Barkin R.L.
      • Webb F.
      • Wilson G.
      Urine drug test interpretation: what do physicians know?.
      The following section reviews commonly prescribed opioids, their metabolic pathways (Figure 1), and their utility in UDT.
      Morphine and codeine are both derived from opium. Codeine is metabolized to morphine and norcodeine. In the urine, all 3 compounds can be detected after codeine ingestion. Morphine is metabolized to 3-morphine-glucuronide and 6-morphine-glucuronide. Hydromorphone has been identified as a minor metabolite of morphine.
      • Cone E.J.
      • Heit H.A.
      • Caplan Y.H.
      • Gourlay D.
      Evidence of morphine metabolism to hydromorphone in pain patients chronically treated with morphine.
      Codeine and hydrocodone metabolism can also produce small amounts of hydrocodone and hydromorphone, respectively, and should not be interpreted as indicators of hydrocodone or hydromorphone ingestion when high concentrations of codeine or hydromorphone are detected in the UDT.
      • Chen Y.L.
      • Hanson G.D.
      • Jiang X.
      • Naidong W.
      Simultaneous determination of hydrocodone and hydromorphone in human plasma by liquid chromatography with tandem mass spectrometric detection.
      • Oyler J.M.
      • Cone E.J.
      • Joseph Jr., R.E.
      • Huestis M.A.
      Identification of hydrocodone in human urine following controlled codeine administration.
      Heroin is rapidly metabolized to 6-monoacetylmorphine (6-MAM), which is further deacetylated to morphine. If heroin use is suspected, one can test for 6-MAM in the urine using a definitive method because the 6-MAM metabolite is specific only to heroin and not morphine or codeine. However, 6-MAM has an extremely short half-life of 36 minutes and is detected only up to 8 hours in the urine after heroin use.
      • Cone E.J.
      • Dickerson S.
      • Paul B.D.
      • Mitchell J.M.
      Forensic drug testing for opiates, V: urine testing for heroin, morphine, and codeine with commercial opiate immunoassays.
      In addition, street heroin may be adulterated with other opioids, such as acetylcodeine, making it difficult to differentiate between heroin, codeine, or morphine use.
      Critical issues in urinalysis of abused substances: report of the Substance-Abuse Testing Committee.
      Oxycodone is frequently prescribed to treat pain and has been shown to have high abuse potential.
      • Cone E.J.
      • Fant R.V.
      • Rohay J.M.
      • et al.
      Oxycodone involvement in drug abuse deaths: a DAWN-based classification scheme applied to an oxycodone postmortem database containing over 1000 cases.
      Oxycodone is metabolized into the active metabolite oxymorphone and moderately active metabolite noroxycodone.
      • Samer C.F.
      • Daali Y.
      • Wagner M.
      • et al.
      Genetic polymorphisms and drug interactions modulating CYP2D6 and CYP3A activities have a major effect on oxycodone analgesic efficacy and safety.
      • Yee D.A.
      • Best B.M.
      • Atayee R.S.
      • Pesce A.J.
      Observations on the urine metabolic ratio of oxymorphone to oxycodone in pain patients.
      About 13% to 19% of the dose is excreted as unchanged drug, 7% to 29% as oxycodone conjugates, 13% to 14% as oxymorphone metabolite, and an unknown amount to noroxycodone.
      • White R.M.
      • Black M.L.
      Pain Management Testing Reference.
      Large variability of metabolic ratio has been published in the literature identifying abnormal metabolite formation when considering ultra-rapid and poor metabolizers of oxycodone to oxymorphone.
      • Yee D.A.
      • Best B.M.
      • Atayee R.S.
      • Pesce A.J.
      Observations on the urine metabolic ratio of oxymorphone to oxycodone in pain patients.
      Methadone is a potent opioid with unique pharmacology; notably, it has a long elimination half-life, which makes it attractive for treatment of chronic pain and dependence on opioids and heroin.
      • Trescot A.M.
      • Datta S.
      • Lee M.
      • Hansen H.
      Opioid pharmacology.
      • Chang K.C.
      • Huang C.L.
      • Liang H.Y.
      • et al.
      Gender-specific differences in susceptibility to low-dose methadone-associated QTc prolongation in patients with heroin dependence.
      • Shaiova L.
      • Berger A.
      • Blinderman C.D.
      • et al.
      Consensus guideline on parenteral methadone use in pain and palliative care.
      About one-third of methadone is excreted unchanged in the urine and is metabolized to an inactive metabolite 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidene (EDDP). Although both methadone and EDDP are present in the urine, many methadone immunoassays detect only the parent compound, methadone. This can be problematic because patients occasionally spike their urine with their methadone prescription to generate a positive result on a UDT.
      • Galloway F.R.
      • Bellet N.F.
      Methadone conversion to EDDP during GC-MS analysis of urine samples.
      There are screening methods for methadone and EDDP and only 1 assay, Immunalysis’s Homogeneous Enzyme Immunoassay (HEIA), tests for both with a cutoff of 300 ng for methadone and 500 ng for EDDP.
      • Methadone/EDDP homogeneous enzyme immunoassay (HEIA™)
      Moreover, testing for EDDP with GC-MS may be necessary with suspected adulteration and in patients who are rapid metabolizers of methadone. A few medications, including verapamil, diphenhydramine, and doxylamine, have been reported to cause false-positive screens for methadone and requiring secondary confirmation.
      • Lichtenwalner M.R.
      • Mencken T.
      • Tully R.
      • Petosa M.
      False-positive immunochemical screen for methadone attributable to metabolites of verapamil.
      • Rogers S.C.
      • Pruitt C.W.
      • Crouch D.J.
      • Caravati E.M.
      Rapid urine drug screens: diphenhydramine and methadone cross-reactivity.
      • Syed H.
      • Som S.
      • Khan N.
      • Faltas W.
      Doxylamine toxicity: seizure, rhabdomyolysis and false positive urine drug screen for methadone.
      Fentanyl transdermal patch is another widely used opioid mainly due to its convenient nonoral route, but it also poses a high risk of serious adverse effects including respiratory depression.
      • Alonso-Zaldivar R.
      FDA renews warning for powerful painkiller patch—the agency says the drug has been misused and wrongly prescribed.
      • US Food and Drug Administration
      Information for Healthcare Professionals: Fentanyl Transdermal System.
      Fentanyl is extensively metabolized to its major inactive metabolite, norfentanyl.
      • Trescot A.M.
      • Datta S.
      • Lee M.
      • Hansen H.
      Opioid pharmacology.
      Fentanyl has been shown to have high intrasubject variability over time and intersubject variability. In patients with pain disorders, the transdermal fentanyl excretion variability may be due to genetic polymorphism of the CYP3A4, skin absorption, and interactions with drugs used concomitantly that interfere with fentanyl metabolism.
      • Cole J.M.
      • Best B.M.
      • Pesce A.J.
      Variability of transdermal fentanyl metabolism and excretion in pain patients.
      Tramadol is a weak opioid agonist that is commonly used for mild pain. It is a prodrug metabolized to an active metabolite O-desmethyltramadol and inactive metabolite nortramadol. Both these metabolites are further metabolized to inactive O-desmethylnortramadol.
      • Grond S.
      • Sablotzki A.
      Clinical pharmacology of tramadol.
      GC-MS, LC-MS/MS, and other procedures to determine tramadol and its metabolites in the urine have been developed.
      • El-Sayed A.A.
      • Mohamed K.M.
      • Nasser A.Y.
      • Button J.
      • Holt D.W.
      Simultaneous determination of tramadol, O-desmethyltramadol and N-desmethyltramadol in human urine by gas chromatography-mass spectrometry.
      Clinical utility of tramadol drug screening may be important for clinicians to be familiar with as the use of tramadol increases.

      Benzodiazepines

      Benzodiazepines are widely prescribed for use as sedatives, hypnotics, anxiolytics, anticonvulsants, and muscle relaxants.
      • Perry P.J.
      • Alexander B.
      • Liskow B.I.
      • DeVane C.L.
      Psychotropic Drug Handbook.
      More than 15 benzodiazepines are commercially available for use in the United States; in addition, large numbers of other benzodiazepines are available in other countries including flunitrazepam, commonly referred to as the “date rape” drug. Because of their sedative properties, benzodiazepines are frequently misused and abused, and chronic use can lead to physiological dependence and addiction.
      Urine drug testing for benzodiazepines is commonly used to check for medication adherence, evaluate abuse/misuse, or identify medications in overdose or emergency situations. Benzodiazepines are secondary to opiates in accidental or intentional overdose situations and are commonly prescribed with other sedating medications.
      • Jann M.
      • Kennedy W.K.
      • Lopez G.
      Benzodiazepines: a major component in unintentional prescription drug overdoses with opioid analgesics.
      Because of the widespread use of benzodiazepines (eg, sedation in the emergency department setting), it is important that clinicians evaluate patient’s medication regimen extensively when evaluating UDT results.
      Interpretation of urine benzodiazepine immunoassays can be complex secondary to benzodiazepine’s metabolic pathway (Figure 2), half-life, potencies, and the inability to differentiate between individual benzodiazepines.
      • elSohly M.A.
      • Feng S.
      • Salamone S.J.
      • Wu R.
      A sensitive GC-MS procedure for the analysis of flunitrazepam and its metabolites in urine.
      • Lee D.C.
      Sedative-hypnotics.
      • Trevor A.J.
      Sedative-hypnotic drugs.
      Chronic use of diazepam, a long half-life agent, can be detected over 30 days in the urine, whereas triazolam, a short half-life drug, may be detected in the urine only for a day.
      • Substance Abuse and Mental Health Services Administration
      Clinical Drug Testing in Primary Care.
      Benzodiazepines with short half-lives or those that are highly lipophilic (eg, alprazolam and diazepam) tend to have the most risk for abuse. Furthermore, there are 2 significant limitations of benzodiazepine immunoassays that may lead to false-negative results: (1) the immunoassay’s inability to detect conjugated metabolites and (2) high cutoff values.
      Figure thumbnail gr2
      Figure 2Basic metabolism of benzodiazepines.
      • elSohly M.A.
      • Feng S.
      • Salamone S.J.
      • Wu R.
      A sensitive GC-MS procedure for the analysis of flunitrazepam and its metabolites in urine.
      • Lee D.C.
      Sedative-hypnotics.
      • Trevor A.J.
      Sedative-hypnotic drugs.
      aFlunitrazepam in not available in the United States.
      Most benzodiazepine immunoassays are designed to detect the free or nonconjugated forms of oxazepam or nordiazepam, which are common metabolites of several benzodiazepines (eg, diazepam, chlordiazepoxide, and temazepam).
      • Lee D.C.
      Sedative-hypnotics.
      However, many benzodiazepines are excreted as glucuronide conjugates (eg, lorazepam and alprazolam) and will not be detected by most immunoassays unless hydrolysis with beta-glucuronidase is performed on the urine before testing.
      • Meatherall R.
      Benzodiazepine screening using EMIT II and TDx: urine hydrolysis pretreatment required.
      • Meatherall R.
      Optimal enzymatic hydrolysis of urinary benzodiazepine conjugates.
      • Meatherall R.C.
      • Fraser A.D.
      CEDIA dau Benzodiazepine screening assay: a reformulation.
      Most laboratories do not use this technique. Clonazepam is another benzodiazepine that may result in a false-negative result because it is primarily reduced to 7-aminoclonazepam and not converted to oxazepam or its conjugate nor does it cross-react well in the immunoassay screen.
      • West R.
      • Pesce A.
      • West C.
      • et al.
      Comparison of clonazepam compliance by measurement of urinary concentration by immunoassay and LC-MS/MS in pain management population.
      Cutoff concentrations of 200 or 300 ng/mL for benzodiazepines were initially established on the basis of standard dosages of older benzodiazepines such as diazepam, oxazepam, and flurazepam dosed between 5 and 20 mg/d.
      • Meatherall R.
      • Fraser A.D.
      Comparison of four immunoassays for the detection of lorazepam in urine.
      Using a cutoff of 200 or 300 ng/mL often is too high for more potent benzodiazepines that are prescribed at lower doses such as lorazepam, alprazolam, and triazolam. Fraser and Meatherall
      • Fraser A.D.
      • Meatherall R.
      Comparative evaluation of five immunoassays for the analysis of alprazolam and triazolam metabolites in urine: effect of lowering the screening and GC-MS cut-off values.
      • Fraser A.D.
      • Meatherall R.
      Improved cross-reactivity to alpha OH triazolam in the BMC CEDIA DAU urine benzodiazepine assay.
      found that lowering the cutoff concentration of alprazolam and triazolam to 100 ng/mL along with enzyme hydrolysis increased positive results. In addition, West et al
      • West R.
      • Pesce A.
      • West C.
      • et al.
      Comparison of clonazepam compliance by measurement of urinary concentration by immunoassay and LC-MS/MS in pain management population.
      recommended lowering the cutoff level to 40 ng/mL to detect clonazepam’s main metabolite 7-aminoclonazepam.
      Despite the high rate of false-negative results, medications that produce false-positive results on the benzodiazepine immunoassays are minimal (Table 4). Sertraline, a commonly prescribed medication for treatment of depression, has widely been reported to cause false-positive results with benzodiazepine immunoassays with rates of 27% to 32% found in 2 retrospective studies.
      • Lum G.
      • Mushlin B.
      • Farney L.
      False-positive rates for the qualitative analysis of urine benzodiazepines and metabolites with the reformulated Abbott Multigent reagents.
      • Nasky K.M.
      • Cowan G.L.
      • Knittel D.R.
      False-positive urine screening for benzodiazepines: an association with sertraline? A two-year retrospective chart analysis.
      Oxaprozin and efavirenz are additional agents that have also been found to interfere with the urine immunoassays.
      • Blank A.
      • Hellstern V.
      • Schuster D.
      • et al.
      Efavirenz treatment and false-positive results in benzodiazepine screening tests.
      • Roder C.S.
      • Heinrich T.
      • Gehrig A.K.
      • Mikus G.
      Misleading results of screening for illicit drugs during efavirenz treatment.
      However, efavirenz’s interference has been found to occur only in the Triage 8 urine drug test and Drug Screen Multi 5 test.
      • Blank A.
      • Hellstern V.
      • Schuster D.
      • et al.
      Efavirenz treatment and false-positive results in benzodiazepine screening tests.
      • Roder C.S.
      • Heinrich T.
      • Gehrig A.K.
      • Mikus G.
      Misleading results of screening for illicit drugs during efavirenz treatment.

      CNS Stimulants

      Amphetamines

      There are an estimated 1.6 million people aged 12 years and older (0.6% of the population ≥12 years) who reported using stimulants for nonmedical uses.
      • Substance Abuse and Mental Health Services Administration
      Behavioral Health Trends in the United States: Results from the 2014 National Survey on Drug Use and Health.
      Among those who reported current use of stimulants, two-third reported abusing prescription stimulants but not methamphetamine. Amphetamines are commonly abused for their euphoric and stimulant effects, and prescription amphetamines have been favored by college students for their supposed “cognitive effects.”
      Amphetamine immunoassays are the screening tests most commonly associated with false-positive results due to the presence of other cross-reacting drugs and substances. It is difficult to develop antibodies that are specific to amphetamine and methamphetamines due to their structures. Methamphetamine also has 2 isomers (d-methamphetamine and l-methamphetamine) that contribute to issues with cross-reactivity and false-positive test results.
      • Eskridge K.D.
      • Guthrie S.K.
      Clinical issues associated with urine testing of substances of abuse.
      Amphetamine assays can detect amphetamines, its isomers (eg, dextroamphetamine), and other amphetamine-type compounds such as methamphetamine, methylenedioxyethylamphetamine, methylnenedioxyamphetamine, and methylenedioxymethylamphetamine as well as other metabolically produced amine-containing compounds.
      Agents that have been commonly linked to false-positive amphetamine results include pseudoephedrine/ephedrine,
      • Stout P.R.
      • Klette K.L.
      • Horn C.K.
      Evaluation of ephedrine, pseudoephedrine and phenylpropanolamine concentrations in human urine samples and a comparison of the specificity of DRI amphetamines and Abuscreen online (KIMS) amphetamines screening immunoassays.
      bupropion,
      • Nixon A.L.
      • Long W.H.
      • Puopolo P.R.
      • Flood J.G.
      Bupropion metabolites produce false-positive urine amphetamine results.
      labetalol,
      • Gilbert R.B.
      • Peng P.I.
      • Wong D.
      A labetalol metabolite with analytical characteristics resembling amphetamines.
      • Yee L.M.
      • Wu D.
      False-positive amphetamine toxicology screen results in three pregnant women using labetalol.
      and ranitidine.
      • Grinstead G.F.
      Ranitidine and high concentrations of phenylpropanolamine cross react in the EMIT monoclonal amphetamine/methamphetamine assay.
      • Liu L.
      • Wheeler S.E.
      • Rymer J.A.
      • et al.
      Ranitidine interference with standard amphetamine immunoassay.
      • Poklis A.
      • Hall K.V.
      • Still J.
      • Binder S.R.
      Ranitidine interference with the monoclonal EMIT d.a.u. amphetamine/methamphetamine immunoassay.
      Bupropion’s chemical structure is similar to those of amphetamines and contributes to the false-positive results.
      • Nixon A.L.
      • Long W.H.
      • Puopolo P.R.
      • Flood J.G.
      Bupropion metabolites produce false-positive urine amphetamine results.
      Metformin has also been linked to false-positive results for amphetamines.
      • Fucci N.
      False positive results for amphetamine in urine of a patient with diabetes mellitus.
      The mechanism is unknown for metformin’s interference, but the importance of confirmatory testing was stressed by one author to avoid negative consequences for patients. Additional medications and products that are not obvious culprits for causing positive results for amphetamines include selegiline and Vick’s Vapor inhalers. Selegiline is metabolized into l-methamphetamine, l-desmethylselegiline, and l-amphetamine that contribute to its cross-interference with amphetamine assays. Selegiline’s metabolites have also been detected in hair up to 4 weeks after a single oral dose.
      • Kronstrand R.
      • Andersson M.C.
      • Ahlner J.
      • Larson G.
      Incorporation of selegiline metabolites into hair after oral selegiline intake.
      Vick’s Vapor Inhalers have been reported to contain 1% to 2.5% d-methamphetamine.
      • Smith M.L.
      • Nichols D.C.
      • Underwood P.
      • et al.
      Methamphetamine and amphetamine isomer concentrations in human urine following controlled Vicks VapoInhaler administration.
      In Smith et al
      • Smith M.L.
      • Nichols D.C.
      • Underwood P.
      • et al.
      Methamphetamine and amphetamine isomer concentrations in human urine following controlled Vicks VapoInhaler administration.
      report, d-methamphetamine and l-methamphetamine were not detected in urine at a lower level of quantification of 10 μg/L after 28 inhalations of Vick’s Vapor inhalers. There were no positive test results for d-methamphetamine or d-amphetamine when GC-MS confirmatory testing was used. l-Methamphetamine was present in most urine specimens at 11 hours after the inhalation but at low concentrations (<250 μg/L). Lisdexamfetamine (Vyvanse) is a prodrug that is inactive before ingestion, which may lead to misconceptions that the drug will not be detected in UDT.
      • Schaeffer T.
      Abuse-deterrent formulations, an evolving technology against the abuse and misuse of opioid analgesics.
      It should be noted that on activation in the gastrointestinal tract, lisdexamfetamine is converted to l-lysine and the active d-amphetamine and will be detected in the urine.
      A popular dietary supplement containing dimethylamylamine (DMAA) also known as methylhexamine and geranium extract has been linked to a false-positive amphetamine screen.
      • Vorce S.P.
      • Holler J.M.
      • Cawrse B.M.
      • Magluilo Jr., J.
      Dimethylamylamine: a drug causing positive immunoassay results for amphetamines.
      DMAA has been marketed under the name OxyElite Pro (among others) for enhancing weight loss and athletic performance. It has been estimated that DMAA is present in more than 200 supplements despite reports of the agent’s association with hemorrhagic strokes and death.
      • Cohen P.A.
      A false sense of security? The U.S. Food and Drug Administration’s framework for evaluating new supplement ingredients.
      • Eliason M.J.
      • Eichner A.
      • Cancio A.
      • Bestervelt L.
      • Adams B.D.
      • Deuster P.A.
      Case reports: death of active duty soldiers following ingestion of dietary supplements containing 1,3-dimethylamylamine (DMAA).
      • Gee P.
      • Tallon C.
      • Long N.
      • Moore G.
      • Boet R.
      • Jackson S.
      Use of recreational drug 1,3 dimethylamylamine (DMAA) [corrected] associated with cerebral hemorrhage.
      In an analysis by the Department of Defense, DMAA was found in 92.3% of the false-positive amphetamine samples that were then confirmed to be negative by GC-MS.
      • Vorce S.P.
      • Holler J.M.
      • Cawrse B.M.
      • Magluilo Jr., J.
      Dimethylamylamine: a drug causing positive immunoassay results for amphetamines.

      Cocaine

      Cocaine is a CNS stimulant extracted from coca leaves.
      • Goldstein R.A.
      • DesLauriers C.
      • Burda A.
      • Johnson-Arbor K.
      Cocaine: history, social implications, and toxicity: a review.
      Similar to amphetamines, cocaine is often abused for its euphoric and stimulant effects. It can also produce anorexia, insomnia, and an increased attention span. Although illegal in the United States, some countries use coca leaves in teas, drinks, and other natural products. Ingestion of these products can cause positive results for cocaine UDT.
      Urine testing for cocaine assesses the presence or absence of cocaine’s primary metabolite, benzoylecgonine. Minimal cross-reactivity exists with drug screens for cocaine.
      • Eskridge K.D.
      • Guthrie S.K.
      Clinical issues associated with urine testing of substances of abuse.
      Although amoxicillin is reported from various Internet sources and review articles to produce false-positive results for cocaine, lack of evidence exists to support this finding.
      • Rapuri S.B.
      • Ramaswamy S.
      • Madaan V.
      • Rasimas J.J.
      • Krahn L.E.
      ‘Weed’out false-positive urine drug screens: table of possible false positives.
      Reisfield et al
      • Reisfield G.M.
      • Haddad J.
      • Wilson G.R.
      • et al.
      Failure of amoxicillin to produce false-positive urine screens for cocaine metabolite.
      tested amoxicillin’s theoretical cross-reactivity for cocaine on 4 different immunoassays and found no false-positive results for cocaine metabolites. In clinical practice, cocaine is available for use as a topical anesthetic in otolaryngology and ophthalmic procedures. Topical and ophthalmic use of cocaine can produce true-positive results for cocaine in the urine.
      • Jacobson D.M.
      • Berg R.
      • Grinstead G.F.
      • Kruse J.R.
      Duration of positive urine for cocaine metabolite after ophthalmic administration: implications for testing patients with suspected Horner syndrome using ophthalmic cocaine.
      However, other anesthetic agents such as benzocaine, lidocaine, procaine, and tetracaine are structurally distinct from cocaine and its metabolites and do not produce false-positive results on UDTs.
      • Dasgupta A.
      Beating Drug Tests and Defending Positive Results: A Toxicologist’s Perspective.

      Phencyclidine

      Phencyclidine (PCP), a dissociative anesthetic, is 1 of the 5 mandated drugs of abuse in the Department of Health and Human Services guidelines for workplace UDT. Although PCP abuse declined in popularity in the 1980s and 1990s, there has been a revival of PCP use in the 2000s especially in combination with other illicit substances. In 2011, the Drug Abuse Warning Network found a 400% increase in emergency room visits for PCP from 2005 to 2011.
      • Substance Abuse and Mental Health Services Administration
      The DAWN Report: Emergency Department Visits Involving Phencyclidine (PCP).
      Frequently, abusers of PCP are dipping or spraying marijuana cigarettes with liquid PCP (“embalming fluid,” “rocket fuel”) often referred to as smoking “wet,” “illy,” or “fry.”
      • Modesto-Lowe V.
      • Petry N.M.
      Recognizing and managing “illy” intoxication.
      • Peters Jr., R.J.
      • Williams M.
      • Ross M.W.
      • Atkinson J.
      • McCurdy S.A.
      The use of fry (embalming fluid and PCP-laced cigarettes or marijuana sticks) among crack cocaine smokers.
      Users of PCP-laced marijuana exhibit violent and aggressive behaviors, severe hallucinations, paranoia, and impaired motor skills.
      • Peters Jr., R.J.
      • Kelder S.H.
      • Meshack A.
      • Yacoubian Jr., G.S.
      • McCrimmons D.
      • Ellis A.
      Beliefs and social norms about cigarettes or marijuana sticks laced with embalming fluid and phencyclidine (PCP): why youth use “Fry”.
      In its pure form, PCP is a white crystalline power (“angel dust”) and is commonly snorted, with effects seen in 2 to 5 minutes. Symptoms of intoxication usually last 4 to 6 hours; however, toxicity with large dosages can persist for 48 hours.
      • Olmedo R.E.
      Phencyclidine and ketamine.
      Detection time of PCP in the urine is approximately 8 days.
      False-positive results for PCP on immunoassays have been reported to occur with agents that are structurally similar to PCP such as tramadol, dextromethorphan, diphenhydramine, and ketamine. Several case reports have shown tramadol’s cross-reactivity to occur during tramadol toxicity, secondary to intentional overdose or misuse of the medication resulting in high tramadol concentrations in the urine.
      • Hull M.J.
      • Griggs D.
      • Knoepp S.M.
      • Smogorzewska A.
      • Nixon A.
      • Flood J.G.
      Postmortem urine immunoassay showing false-positive phencyclidine reactivity in a case of fatal tramadol overdose.
      • Ly B.T.
      • Thornton S.L.
      • Buono C.
      • Stone J.A.
      • Wu A.H.
      False-positive urine phencyclidine immunoassay screen result caused by interference by tramadol and its metabolites.
      Rengarajan and Mullins
      • Rengarajan A.
      • Mullins M.E.
      How often do false-positive phencyclidine urine screens occur with use of common medications?.